Motor List
This page will contain the motor lists for the beamlines
The tables below show the various motors accessible via spec for the optics of ID14-4 in OH2 and the experiment in EH4. Links go from the motor names to the pages describing the optical element (where it exists).
1. The Optics Hutch
The first table refers to the motors used to move the optical elements in OH2. The motors are listed inarbitrary order.
| Motor Name in spec ('oh2' on artemis) |
Function |
Callibrated |
Useful Increments/values |
Comments |
| Translation of the monochromator Si-xtals to move 111 in (tmainy=22.74) or to move the 311 in (tmainy=79.08) |
|
Keep fixed on 22.74 |
Don't use |
|
| tmainz |
? |
|
Don't use |
|
| qga |
Piezo crystal for r2theta or r2gamma |
- |
- |
Used by feedback-system |
| qgb |
Piezo crystal for r2gamma or r2theta |
- |
- |
Used by feedback-system |
| Slits before mono, front blade |
mm |
0.1mm |
Used to reduce heatload mono |
|
| Slits before mono, back blade |
mm |
0.1mm |
Used to reduce heatload mono |
|
| Slits before mono, down blade |
mm |
0.1mm |
Used to reduce heatload mono |
|
| Slits before mono, up blade |
mm |
0.1mm |
Used to reduce heatload mono |
|
| Pseudo-motors for the slits before the mono: vertical gap and offset |
mm |
Good value is 2 mm |
Close ss1vg to 0.3 and do a dscan on ss1vo –1.5 1.5 20 1 with plotselect I4 to determine best position |
|
| Pseudo-motors for the slits before the mono: horizontal gap and offset |
mm |
Good value is 3 mm |
Close ss1hg to 0.3 and do a dscan on ss1ho –2 2 20 1 with plotselect I4 to determine best position. |
|
| Vertical positioning legs for the toroidal mirror: mvb back of mirror; mvin closest to storage ring at mirror end; mvout furthest from ring at mirror end |
|
0.01mm |
|
|
| Pseudo-motor for mirror height |
mm |
0.04mm |
Only motor of mirror you could try to move. Use a maximum of 3x40 mu steps |
|
| Pseudo mtor of mvin/mvout giving a roll around the X-rays for the mirror |
radians |
To be installed and tested |
||
| Pseudo mtor of mvf/mvin/mvout giving the glancing angle of X-rays to the mirror |
radians |
To be installed and tested |
||
| Translation horizontally of the mirror |
mm |
Rarely moved |
||
| Pseudo-motor of myawf/myawb for the yaw of the mirror |
radians |
0.01 |
Tweaked once a run or so; focus very sensitive to yaw position |
|
| Pseudo-motor of myawf/myawb for the horizontal translation of the mirror |
mm |
To be installed and tested |
||
| atten1; atten2; atten3 |
Three rows of attenuator |
none |
Use gui if you like to use the attenuators in the white beam. Currently disabled |