MXCuBE troubleshooting
General
First try reloading the MXCuBE 3 web interface (shift-F5) - this fixes most problems.
If this does not solve the problem. Kill and start the MXCuBE v3 server.
From a remote session, in NX, make sure you have started the ID23EH2 script on the desktop
From the green sybil window:
stop-mxcube
then
mxcube
Keep in mind that you must leave this process running -if you kill it or you kill your NX session, it must be started again in order for mxcube3 to work
Once you have restarted mxcube3, you will need to refresh your browser (Control-R or shift-F5 or "Reload" from the menu bar)
If this does not work, trying a different browser can help in some cases.
MXCube server will not start
Try restarting the transfocator device server:
on sybil, type:
supervisorctl
then
restart TRANSFOCATORS:tf1
restart TRANSFOCATORS:tf2
and then try to restart the mxcube server again
Safety shutter cannot be opened/blank images
see "Blank Images" section -- it is probably the RV9 valve.
Still problems e.g. moving motors?
Click at left of MXCUBE3 to show all motors -- are any red? if so, move them with the up or down arrow to turn them green